Мы Вам весьма благодарны за обращение! Мы перезвоним в рабочее время.
Rotating Anode X-ray Tube Assembly
Rotating anode X-ray tube assembly for the
purpose of general diagnostic X-ray
procedures.
Specially processed Rhenium-tungsten faced
molybdenum target of 74 mm diameter.
This tube has foci 1.2 and 0.6, and is available
for a maximum tube voltage 150 kV.
Accommodated with IEC60526 type
high-voltage cable receptacles.
General Data
IEC Classification (IEC60601-1:2005) ......................................... Class Ⅰ ME EQUIPMENT
Electrical:
Circuit:
High Voltage Generator ............................................ Constant Potential High-Voltage Generator
Grounding .......................................................................................................... Center-grounded
Nominal X-ray Tube Voltage (IEC60613:2010):
Radiographic ...................................................................................................................... 150 kV
Nominal Focal Spot Value (IEC60336:2005):
Large Focus ............................................................................................................................. 1.2
Small Focus ............................................................................................................................. 0.6
Nominal Anode Input Power (at 0.1s): (See rating charts)
60Hz 50 Hz
Large Focus ......................................................................................................... 50 kW 46 kW
Small Focus ......................................................................................................... 22 kW 20 kW
Nominal Radiographic Anode Input Power (IEC60613:2010):
60 Hz 50 Hz
Large Focus ......................................................................................................... 50 kW 46 kW
Small Focus ......................................................................................................... 21 kW 19 kW
E7843X
Motor Ratings:
Stator: XS-BA
Starting Running
Driven Frequency [Hz] 50 60 50 60
Input Power [W] 600 2000 520 1700 150 130
Voltage 1) 3) [V] 120 220 120 220 50-60 50-60
Current 2) [A] 3.5~5.5 6.7~8.4 3.5~5.0 5.8~7.5 1.7~2.4 1.4~2.1
Min. Speed Up 4) [s] 2.2 0.8 2.2 0.8 - -
Capacitor [µF] 43 30 43 30
Note 1) Applied voltage between common and main terminal.
2) Common current.
3) The every applied voltage must be never exceeded 110% of the above specification.
4) The speed-up time is allowed up to 110% of the above specification.
Anode Speed:
60 Hz ............................................................................................................ Minimum 3200 min-1
50 Hz ............................................................................................................ Minimum 2700 min-1
Stator Resistance:
Common-Main Winding ....................................................................................................... 18.0 Ω
Common-Auxiliary Winding .................................................................................................... 47.5 Ω
Resistance between Housing and Low Voltage Terminals ........................................ Minimum 2 MΩ
Normal Operating Range of the Housing Temperature ..................................................... 16 ~ 75 ℃
Mode of Operation ............................................................................................................. Intermittent
Mechanical:
Dimensions .................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 479 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Anode Angle ............................................................................................................... 12 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Filtration:
Permanent Filtration ................................................................ 1.3 mm Al / 75 kV Available Additional Filter combination (0.4 - 1.5 mm) ..................... IEC60522:1999
Maximum 2.8 mm Al / 75 kV
Radiation Protection (In accordance with IEC60601-1-3:2008):
Leakage Technique Factor .................................................................................. 150 kV, 3.4 mA
X-ray Coverage ................................................................................ 430 × 430 mm at SID 1000 mm
Weight (Approx.) ........................................................................................................................ 16 kg
High Voltage Receptacle ........................... To meet requirements of IEC60526 Corrigendum1:2010
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number ................................................................................................ XH-12
Absolute Maximum and Minimum Ratings
(At any time, these values must not be exceeded.)
Maximum X-ray Tube Voltage (IEC60613:2010):
Radiographic ...................................................................................................................... 150 kV
Between Anode (or Cathode) and Ground ................................................................................ 75 kV
Minimum X-ray Tube Voltage ................................................................................................... 40 kV
Maximum X-ray Tube Current (IEC60613:2010) .................................................... See rating charts
Large Focus ...................................................................................................................... 760 mA
Small Focus ...................................................................................................................... 370 mA
Maximum Filament Current:
Large Focus .......................................................................................................................... 5.0 A
Small Focus .......................................................................................................................... 4.7 A
Filament Voltage:
Large Focus (At maximum filament current 5.0 A) ................................................... 12.4 ~ 17.0 V
Small Focus (At maximum filament current 4.7 A) ....................................................... 6.5 ~ 8.9 V
Filament Frequency Limits ................................................................................................ 0 ~ 25 kHz
Continuous Anode Input Power (IEC60613:2010) ........................................................ 60 W (85 HU/s)
Thermal Characteristics:
Anode Heat Content ......................................................................................... 111 kJ (150 kHU)
Maximum Anode Heat Dissipation ................................................................... 475 W (667 HU/s)
X-ray Tube Assembly Heat Content ................................................................ 900 kJ (1250 kHU)
Nominal Continuous Input Power (IEC60613:2010):
Without Air-circulator ................................................................................ 180 W (15 kHU/min)
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
- 3 -E7843X
The information about EMC conformity (IEC60601-1-2:2007)
As the result of reviewing the test requirements in the following Tables, an X-ray tube assembly
as a component of X-ray system or apparatus does not fall within the scope of the EMC directive of
IEC60601-1-2:2007. Therefore, X-ray system manufacturer shall conduct the required test and
evaluate EMC.
CAUTION: The diagnostic X-ray system can only allow to use the designated cables and
components. If the different cables or components are used, they may deteriorate
the performance of electromagnetic emissions and immunity.
The following tables are the copies of statements in IEC60601-1-2:2007
Table 1 - Guidance and manufacturer's declaration - electromagnetic emissions -
for all EQUIPMENT and SYSTEMS (see 6.8.3.201 a) 3))
Guidance and manufacturer's declaration - electromagnetic emissions
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used
in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The [EQUIPMENT or SYSTEM] uses RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Group 2 The [EQUIPMENT or SYSTEM] must emit
electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be
affected.
RF emissions
CISPR 11
Class [A or B]
Harmonic
emissions
IEC61000-3-2
Class [A, B, C, D, or
Not applicable]
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
[Complies or Not
applicable]
[See 6.8.3.201 a] 3)
and Figure 201
The [EQUIPMENT or SYSTEM] is suitable for use in
all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
[See 6.8.3.201 a] 3)
and Figure 201
The [EQUIPMENT or SYSTEM] is suitable for use in
all establishments other than domestic and these
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
RF emissions
CISPR 14-1
Complies The [EQUIPMENT] is not suitable for interconnection
with other equipment.
RF emissions
CISPR 15
Complies The [EQUIPMENT] is not suitable for interconnection
with other equipment.
- 4 -E7843X
Table 2 - Guidance and manufacturer's declaration - electromagnetic immunity -
for all EQUIPMENT and SYSTEMS (see 6.8.3.201 a) 6))
Guidance and manufacturer's declaration - electromagnetic immunity
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used
in such an environment.
Immunity test IEC60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC61000-4-5
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC61000-4-11
<5 % UT
(>95 % dip in UT )
for 0.5 cycle
40 % UT
(60 % dip in UT )
for 5 cycle
70 % UT
(30 % dip in UT )
for 25 cycle
<5 % UT
(>95 % dip in UT )
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If that user of the
[EQUIPMENT or SYSTEM] requires
continued operation during power
mains interruptions, it is recommended
that the [EQUIPMENT or SYSTEM] be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC61000-4-8
3 A/m Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note: UT is the A.C mains voltage prior to application of the test level.
- 5 -E7843X
Table 3 - Guidance and manufacturer's declaration - electromagnetic immunity -
for LIFE-SUPPORTING EQUIPMENT and SYSTEM (see 6.8.3.201 b))
Guidance and manufacturer's declaration - electromagnetic immunity -
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used
such an environment.
Immunity test IEC60601
test level
Compliance
level Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Portable and mobile RF communications
equipment should be used no closer to any
part of the [EQUIPMENT or SYSTEM],
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
5.3
[V1] V P
d ]
=
[
V
1
Radiated RF
IEC61000-4-3
3 Vrms
150 kHz to 80 MHz
outside ISM bandsa
10 Vrms
150 kHz to 80 MHz
in ISM bandsa
10 V/m
80 MHz to 2.5GHz [V2] V P
d ]
=
[ 12
V
2
d
=
12 [E1] V/m P
[ 1
E ]
80 MHz to 800 MHz
d
=
23 [ 1
E ]
P
800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,c should be less than the
compliance level in each frequency ranged
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Continued to the next page.
- 6 -E7843X
Note: 1. At 80 MHz and 800 MHz, the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, object and people.
a. The ISM (industrial, scientific and medical) bands between 150kHz and 80MHz are 6,765MHz
to 6,795MHz; 13,553MHz to 13,567MHz; 26,957MHz to 27,283MHz; and 40.66MHz to
40.70MHz.
b. The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the
frequency range 80MHz to 2.5GHz are intended to decrease the likelihood that mobile/portable
communications equipment could cause interference if it is inadvertently brought into patient
areas. For this reason, an additional factor of 10/3 is used in calculating the recommended
separation distance for transmitters in these frequency ranges.
c. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the [EQUIPMENT or SYSTEM] is used
exceeds the applicable RF compliance level above, the [EQUIPMENT or SYSTEM] should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the [EQUIPMENT or SYSTEM].
d. Over the frequency range 150kHz to 80MHz, field strengths should be less [V1] V/m.
- 7 -E7843X
Table 4 - Guidance and manufacturer's declaration - electromagnetic immunity -
for EQUIPMENT and SYSTEM that are not LIFE-SUPPORTING (see 6.8.3.201 b))
Guidance and manufacturer's declaration - electromagnetic immunity -
The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment specified
below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used
in such an environment.
Immunity test IEC60601 test
level
Compliance
level Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the [EQUIPMENT or SYSTEM],
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
5.3
d ]
=
[
[V1] V P
V
1
d
=
5.3 [E1] V/m P
[ 1
E ]
80 MHz to 800 MHz
d
=
[ 1
7 800 MHz to 2.5 GHz
E ]
P
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the
compliance level in each frequency rangeb
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note: 1 At 80MHz and 800MHz, the higher frequency range applies.
2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, object and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the [EQUIPMENT or SYSTEM] is used
exceeds the applicable RF compliance level above, the [EQUIPMENT or SYSTEM] should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the [EQUIPMENT or SYSTEM].
b. Over the frequency range 150kHz to 80MHz, field strengths should be less [V1] V/m.
- 8 -E7843X
Warning
Warning to Interface with X-ray Generator
1. Housing Rupture
Never input over rated power to X-ray tube assembly.
If the input power exceeds tube specification, it causes the over temperature of anode, insert tube
glass shatter and ultimately the following serious problems due to generating over-pressure by oil
vaporization inside housing assembly.
In such a critical condition causing housing rupture by over load, the safety thermal switch can not
protect X-ray tube even if it works.
* Housing sealing parts rupture.
* Human injury including burns due to hot oil escape.
* Fire accident due to flaming anode target.
The X-ray generator should have a protective function which manages input power to be within
tube specification.
2. Electric Shock
protective earth.
To avoid the risk of electric shock, this equipment must only be connected to a supply with
- 9 -E7843X
Cautions
Caution to Interface with X-ray Generator
1. Over Rating
X-ray tube assembly can be broken with applying just one over rated shot.
Please read the technical data sheets carefully and follow the instructions.
2. Permanent Filtration
The total filtration and the distance between X-ray focal spot and human body are regulated legally.
They should be complied with the regulation.
3. Safety Thermal Switch
X-ray tube assembly has safety thermal switch to prohibit further input power when the tube
housing reaches to the temperature of switch-open.
The switch should be hooked up with the X-ray generator which control output power to X-ray tube
assembly.
The switch is not recommended connecting stator coil in series circuit.
Even if the switch works, never turn the system power off. The cooling unit should be activated if
used with the system.
4. Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or
failure. If the serious problems caused by the above risk is expected, you are requested to have a
contingency plan to avoid such a case.
5. New Application
If you use the product with new application not to be mentioned in this specification or with
different type of X-ray generator, please contact to us for confirming its availability.
- 10 -E7843X
Caution for Installation, Adjustment and Maintenance
1. Qualified Persons
tube assembly.
Only qualified persons who have technical training and professional knowledge can handle X-ray
2. Fragile Glass
X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or
pulsed shock over 19.6m/s2 (2G).
Careful handling is required to treat or transport.
3. Ground Terminal
X-ray tube assembly has ground terminal. Ground cable should be connected.
4. High Voltage
All X-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the
high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled and high
voltage capacitors / cables discharged.
5. High Voltage Plug
High voltage plug should be cleaned up and free from any physical damages. Silicon compound
application is required for high voltage stability.
6. Rotating Anode
(1) Test rotation of the anode with the large focus filament at the preheat condition. At this time, do
not apply the tube voltage.
(2) When the hand switch "READY" button is pressed, the filament will become brighter and the
anode target will start rotating. The anode for normal speed will reach the predetermined
number of revolution in 0.4 to 1 sec., and it will continue rotation for some time after the stator
drive input is cut off.
The anode for high speed shall be automatically braked to reduced speed and continue
coasting after the drive input to the stator is cut off.
- 11 -E7843X
(3) Ensure that the anode target rotates in the proper direction.
X-ray tube should always rotate counter-clockwise, viewing from the cathode side.
- Confirm direction of rotation -
(4) If the inside of any tube unit cannot be observed (for example, the tube unit with a beryllium
window, etc.), check the tube unit to ensure again that the terminal numbers of the drive circuit
of the low-voltage terminals are properly connected.
7. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.
8. Protective Cover
X-ray tube assembly is not allowed to be used without the protective cover attached.
9. Handling
Appropriate jig or tools are required for tube installation to avoid physical damages.
10. Returning Tube
X-ray tube assembly should be repackaged with the original material when it is returned back for
quality examination in our factory. Be careful to put the tube upside cathode. If the packaging is not
proper, the tube may not be correctly examined. CETD does not warrant it if returned X-ray tube
assembly is in damage at receiving.
- 12 -E7843X
Caution in Rupture Disc
The Rupture Disc is mounted on anode side housing wall.
When external cover is removed, it is visibly recognized as labeled with label 3 in Caution Labels
page.
Followings are important notice of the Rupture Disc:
Function:
When housing inner pressure reaches to the working pressure of the Rupture Disc, the Rupture
Disc collapses and vaporized oil and hot oil flow through the collapsed Rupture Disc.
The external cover can protect the initial impact of oil flow, however, it can not protect oil drop or oil
leakage.
Caution:
1. 2. Never touch the Rupture Disc or remove the label and the screws.
Otherwise, it could cause collapse of Rupture Disc.
Please note the Rupture Disc is not the device to protect X-ray tube from destruction when over
rated power input but just to reduce the impact of housing rupture.
- 13 -E7843X
Caution in Operation
1. X-ray Radiation
X-ray tube assembly should have the beam limiting equipment mounted on the X-ray port to
protect unnecessary radiation.
2. Dielectric Oil
X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for
human health, if it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.
3. Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas.
4. Disposal
X-ray tube housing is lined with lead to protect unnecessary radiation.
The lead powder or vapor is harmful for human health.
Dispose the scrapped products according to the requirement of local regulation.
5. Adjust the Tube Current
Depending on the operating conditions, the filament characteristics might be changed.
This change might lead to the over rate exposure to X-ray tube assembly.
To prevent the X-ray tube assembly from being damaged, adjust the tube current regularly.
Besides when "ROTANODE" has arcing problem in a long time use, adjustment of the tube current
is required.
6. X-ray Tube Housing Temperature
Do not touch on X-ray tube housing surface just after operation due to high temperature.
Stay X-ray tube to be cooled.
7. Operating Limits
Prior to usage, please confirm the environmental condition is within the operating limits.
8. Any Malfunction
Please contact to CETD immediately, if any malfunction of the X-ray tube assembly is noticed.
Optional Parts
Following are additional Aluminium Filters as optional parts.
KITX-104 : A set of 0.4mm Aluminium Filters.
KITX-105 : A set of 0.5mm Aluminium Filters.
KITX-106 : A set of 0.6mm Aluminium Filters.
- 14 -E7843X
Caution Labels
These labels are caution label to notify the user of the following point.
1. "Always have service personnel install or remove the high voltage cable."
2. "Housing end caps must always be mounted to the tube housing assembly properly and correctly
for X-ray protection and safety."
3. "Do not impact on this fragile face."
- 15 -E7843X
Definition Symbol Marks
CE MARKING OF CONFORMITY
SERIAL NUMBER
MANUFACTURER
DATE OF MANUFACTURE
AUTHORIZED REPRESENTATIVE IN THE EUROPEAN COMMUNITY
PROTECTIVE EARTH
EARTH
CAUTION
OPERATING INSTRUCTIONS
- 16 -E7843X
To Keep &l